Stat Profile pHOx Series
- Deskripsi
- Spesifikasi
- Video Produk
- Support
- Download Browsure
Using a unique combination of advanced optical and electrode technology, the Stat Profile pHOx analyzer series offer essential blood gas and critical care test menus in the smallest analyzers currently available. The three-test Stat Profile pHOx Basic offers the standard blood gas profile of pH, PCO2, and PO2. The six-test Stat Profile pHOx menu adds measured hematocrit, hemoglobin, and oxygen saturation to provide a complete picture of lung function and oxygen status.
Stat Profile pHOx Plus adds glucose, sodium, potassium, and a choice of chloride or ionized calcium, resulting in an ideal critical care menu in a small, space-saving instrument. Stat Profile pHOx Plus C offers both chloride and ionized calcium. Stat Profile pHOx Plus L menu includes lactate.
Automatic Blood Gas/Oxymetry Analyzer
Measured Parameters : pH, pCO2, pO2, SO2%, Hb, Hct, Na, K, Cl, iCa, Glu, and Lactate.
Measured / Range | : | pH | : 6.50 - 8.00 | Na | : 80 - 200 mmol/L | ||
pCO2 | : 3.0 - 200 mmHg | K | : 1 - 20 mmol/L | ||||
pO2. | : 0 - 800 mmHg | Cl | : 50 - 200 mmol/L | ||||
SO2% | : 30 - 100 % | iCa | : 0.1 - 2.7 mmol/L | ||||
Hb | : 4.0 - 24 g/dL | Glu | : 15 - 500 mg/dL | ||||
Hct | : 12 % - 70 % | Lac | : 0.3 - 20 mmol/L | ||||
Barometric : 450 - 800 ± 2 mmHg |
Measuring Technology | : | Ion Selective Electrodes ( pH, Na, K, Cl, iCa ) |
Enzyme / Amperometric ( Glucose & Lactate ) | ||
Severinghause (pCO2) | ||
Amperometry ( pO2 ) | ||
Conductivity ( Hematocrit ) | ||
Optical, Reflectance Photometry ( Oxygen Saturation ) | ||
Conductivity/Na+ correction ( Hematocrit ) | ||
Multi Wavelength reflectance/conductivity correction (Hemoglobin) | ||
Acceptable Samples | : | Whole Blood (heparinized), arterial, mixed venous, capillary |
Sample Throughput | : | 40 samples per hour |
Sample Size | : | 125 microliters whole blood, Full Panel |
60 microliters whole blood, for Blood Gases/pH | ||
Acceptable Container | : | Syringes, Capillaries, Vacuum Tubes |
Temperature Control | : | Thermostatting 37°C ±0.1°C |
Calculated | : | Total CO2, Oxygen Content, A, A-aDO2, a/A, R (pO2/FIO2) Actual - |
Bicarbonate, Standard Bicarbonate, Base Excess ( ecf ) , Anion Gap | ||
Base Excess (blood), pH, pCO2, pO2 Values corrected to - | ||
patient temperatures, | ||
Data Capture | : | 8,4 " color LCD touch screen display multilingual, |
On Board QC Package | ||
On Board Patient Storage n=96 | ||
Universal ASTM Interface | ||
Auto Flag On Abnormal & Panic | ||
Bar Code Compatible | ||
52 Column Thermal dot matrix printer | ||
Automatic Reagent Management System | ||
Optional Patient Data Management (PDM) interface | ||
with the LIS | ||
Interfaces | : | USB, RS232 serial communication |
ASTM standard | ||
Calibration | : | User selectable, fully automatic two-point calibration every 2,4, or 6 hours, |
User selectable, single-point calibration every 30/45 minutes or with each sample | ||
Stand by mode with automatic wake-up can be used to conserved reagent | ||
Manual calibration initiated at any time | ||
Dimentions and weight | : | 30.5(W) X 38.1(D) X 38.1(H) cm, 8.19 Kg |
Electricity | : | Automatic Blood Gas/Oxymetry Analyzer |
AC 90 - 264 VAC 50/60 Hz, 130 W |
Silahkan hubungi kontak dibawah ini untuk pertanyaan seputar produk
maupun pemesanan melalui panggilan telephone maupun chat whatsapp