StatStrip and StatStrip Xpress Hospital Glucose Meters
- Deskripsi
- Spesifikasi
- Video Produk
- Support
- Download Browsure
This device [StatStrip Glucose] provides an important public health resource for critically ill hospitalized patients, who often have conditions or are taking medications that can cause incorrect blood glucose reading. It is important for manufacturers of glucose meters used in hospitals to design and test their devices for use in all hospitalized patients.
The only glucose meters cleared by US FDA for use with all patients, including critically ill
Because of the unacceptably high rate of adverse patient events, including deaths, linked to the point-of-care (POC) use of glucose meters within hospitals, there is a recognized need for more accuracy and better performance standards for POC glucose testing in hospitals, particularly in intensive care settings.
StatStrip is designed specifically to measure and correct for clinical interferences that can be present in critically ill patients.
Proven safety and accuracy
- In one study of over 17,300 capillary results from critically ill patients that were compared to lab results
- In another study of 1,698 arterial and venous results from critically ill patients with over 257 medical conditions were compared to lab results
- Improved accuracy results in less insulin misdosing and better outcomes for critically ill patients
GLUCOSE KETONE HOSPITAL METER
(Connected with LIS/HIS)
- Interferences Eliminated :
- Hematocrit
- Maltose
- Galactose
- Ascorbic Acid
- Uric Acid
- Paracetamol (Acetaminopen)
- Bilirubin
- Xyloce
- Oxygen
- Connectivty :
- Meter Data Output RJ-45 Ethernet Port
- Protocol TCP/IP ethernet 10 Mbit
- Standard POCT1-A Compliant
- Setup Program NovaNet, Web-Based Instrument Manager Software
- Additional Features
- Touch Screen
- Built in Barcode reader
- Accepting patient ID or Accention Number
- Use of operator Password (Lock out)
- Connect to All LIS (Laboratory information System) or HIS (Hospital information System)
- Equipped Carrying Case
- Units of measure mmol/L atau mg/dL
- Strip of Glucose and Ketone
- Test Measured Blood Glucose, Hematocrit corrected
- Test Reported Glucose
- Test Time 6 seconds
- Test Strip Volume 1,2 µL
- Test Methodology Electrochemostry
- Sample Types & Operating Modes
- Whole Blood Arterial, Venous, Capillary
- Measurement Ranges
- Glucose 0,5 – 33,3 mmol/L (10 – 600 mg/dL)
- Operating Ranges
- Hematocrit 20 % - 65 %
- Humidity 10 % - 90 %
- Meter Size
- Weight 360 g (0,8 libs)
- Ukuran 153 mm x 82,5 mm x 46 mm (6.0 in x 3.25 in x 1.8 in)
- Battery Type 7 V Li Polymer Rechargeable Battery
- Features Rechargeable/Replaceable
- Life 8 Hours in use (approximately 40 tests w/ Barcode
Scans / 12–24 hours standby
- Test Strip & QC Stability
- Glucose Strip 30 months from date of manufacture
6 months open-vial stability
- Quality Control 3 months open-vial stability
- Other ingredients
- QC consists of 3 levels, Level 1, 2 and 3 (Low, Normal dan High)
- Linearity consists of 5 levels
- Data Storage
- Patient test 1000 Test
- QC test 200 Test
- Users 4000 users
- Docking Station
- Desk or Wall mount
- 3 LED Indicators Lights showing :
- Connection status
- Transmit / Receive data
- Charging status
- Extra battery slot included for recharging and storage of spare battery
Silahkan hubungi kontak dibawah ini untuk pertanyaan seputar produk
maupun pemesanan melalui panggilan telephone maupun chat whatsapp